Quantitative Analysis and Efficiency Study of PSD Methods for a LaBr3:Ce Detector

Ming Zeng,Jirong Cang,Zhi Zeng,Xiaoguang Yue,Jianping Cheng,Yinong Liu,Hao Ma,Junli Li
DOI: https://doi.org/10.1016/j.nima.2015.12.045
IF: 1.335
2016-01-01
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Abstract:The LaBr3:Ce scintillator has been widely studied for nuclear spectroscopy because of its optimal energy resolution (<3%@ 662 keV) and time resolution (similar to 300 ps). Despite these promising properties, the intrinsic radiation background of LaBr3:Ce is a critical issue, and pulse shape discrimination (PSD) has been shown to be an efficient potential method to suppress the alpha background from the Ac-227. In this paper, the charge comparison method (CCM) for alpha and gamma discrimination in LaBr3:Ce is quantitatively analysed and compared with two other typical PSD methods using digital pulse processing. The algorithm parameters and discrimination efficiency are calculated for each method. Moreover, for the CCM, the correlation between the CCM feature value distribution and the total charge (energy) is studied, and a fitting equation for the correlation is inferred and experimentally verified. Using the equations, an energy-dependent threshold can be chosen to optimize the discrimination efficiency. Additionally, the experimental results show a potential application in low-activity high-energy gamma measurement by suppressing the alpha background. (C) 2016 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?