Design and Implementation of a Peak Picking System for MVT Digitizer

H. Lao,Q. Liu,Y. Ling,W. Hu,Y. Liu,B. Ge,Y. Li,H. Yu,Q. Xie
DOI: https://doi.org/10.1109/nssmicrtsd49126.2023.10338509
2023-01-01
Abstract:Multi-Voltage Threshold (MVT) digitizer is a well-known integrated digital signal processing method for new radiation detection technology in Positron Emission Tomography (PET) imaging. However, MVT may face challenges in selecting appropriate threshold voltages for scintillation pulses with a wide range of energy spans, which can limit the PET signal reconstruction performance in terms of energy and time. To address this issue, we propose incorporating peak information into the MVT method. Traditional methods of obtaining peaks using discrete components result in unacceptable high costs and power consumption, particularly in practical PET multi-channel applications. In this paper, we propose a novel 36-channel peak acquisition system chip for MVT digitizers using multi-channel multiplexing technology. We also preliminarily evaluate the multiplexed Analog to Digital Converter (ADC) performance of our design. This provides a basis for the implementation of Peak Picking-MVT in clinical PET detectors. This is expected to have a positive impact on other scintillator-based applications.
What problem does this paper attempt to address?