Long-term Follow-Up of AAISafeR Pacing Mode.

Jihong Guo
2009-01-01
Abstract:Objective To study and compare the long-term results in patients implanted with Symphony 2450.Methods A long-term follow-up survey was carried out in 40 cases by clinical examination,electrocardiography,echocardiography and pacemaker programme.Left atrial diameter(LAD),left ventricular end-diastole diameter(LVEDD)were observed and the cumulative percent ventricular pacing (Cum%VP) during pacemaker programing was calculated.Result The mean follow-up time was18±4.6(12-24)months.The Cum%VP was 23.5%±3.5%.The LAD and LVEDD statistically decreased after pacemaker implantation compared with that before implantation.Atrial arrhythmia was obviously less during follow-up period.Conclusion The mode of AAISafeR could diminish the Cum%VP and the incidence of atrial arrhythmia efficiently.