Study of Flow Injection-Negatively Pressurized Sampling Equipment in Capillary Electrophoresis

Jianwei Wang,Zai Hai-yun,Zuanguang Chen,Peng Cai,Jinhan Mo
2006-01-01
Abstract:A flow injection-negatively pressurized sampling(FI-NPS) equipment was designed.The sample was carried to the injection valve by a peristaltic pump and carried to the inlet of separation capillary by buffer solution,then sample injection into capillary was performed through the negative pressure set in the outlet of capillary.Duration of injection was controlled by a home-made precise time circuit.With the optimization of the injection parameters,better repeatability and simple operation have been achieved.The RSD values of peak area and migration time for two selected cations(Na~(+) and K~(+)) were within the range of 0.7%~2.7% using this injection method.A rapid and efficient on-line analysis could be carried out in the CE system coupling with this equipment and contactless conductivity detector(CCD).In the application to the inorganic cation separation,satisfactory results that the performance of sampling and baseline separation of six cations has been achieved in less than 10 min were observed.