Sequential injection sample introduction microfluidic-chip based capillary electrophoresis system

Qun Fang,Fu-Ren Wang,Shi-Li Wang,Shu-Sheng Liu,Shu-Kun Xu,Zhao-Lun Fang
DOI: https://doi.org/10.1016/S0003-2670(99)00183-X
IF: 6.911
1999-01-01
Analytica Chimica Acta
Abstract:A sequential injection micro-sample introduction system was coupled to a microfluidic-chip based capillary electrophoresis system through a split–flow sampling interface integrated on the micro-chip. The microfluidic system measured 20×70×3mm in dimension, and was produced using a non-lithographic approach with components readily available in the analytical laboratory. In the H-configuration channel design the horizontal separation channel was a 75μm I.D.×60mm quartz capillary, with two vertical side arms produced from plastic tubing. The conduits were embedded in silicon elastomer with a planar glass base. Sequential introduction of a series of samples with about 2.5% carryover was achieved at 48h−1 throughput with samples containing a mixture of fluorescein isothiocyanate (FITC)-labeled amino acids using SI sample volumes of 3.3μl and carrier flow-rate of 2.0mlmin−1. Baseline separation was achieved for FITC-labeled arginine, phenylalanine, glycine and FITC (laser induced fluorescence detection) in sodium tetraborate buffer (pH 9.2) within 8–80s, at separation lengths of 25–35mm and electrical field strengths of 250–1500Vcm−1, with plate heights in the 0.7–3μm range.