Stack-Layer Dual-Frequency Ultrasound Array with Ground Shielding for Super-Harmonic Imaging

Jianguo Ma,Tieming Liu,Lijun Xu
DOI: https://doi.org/10.1109/tim.2023.3332397
IF: 5.6
2024-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:Contrast-enhanced harmonic imaging has shown unique advantages in identifying micro-vasculature for tumor characterization. Recent studies indicate that the 3rd to 4th harmonics are the optimal ranges to detect microbubble contrast agents in microvessels. However, it is challenging for conventional piezoelectrical transducers to detect high-order harmonics due to their limited bandwidth. This paper describes a stacked dual-frequency array for microvessel detection, which comprises 32 transmitting (TX) elements at 2 MHz and 96 receiving (RX) elements at 7 MHz. Design with ground shielding at the front surface ensures the safety and low noise of the transducer. Real-time contrast imaging was performed with vessel phantom inside water or porcine tissue. The contrast to noise ratio in water was 23 dB, whereas the contrast to tissue ratio in porcine was 16 dB in the two experiments. These results suggest that the array has a potential for detecting microvessels with high fidelity.
What problem does this paper attempt to address?