Design and Fabrication of a Novel Dual-Frequency Confocal Ultrasound Transducer for Microvessels Super-Harmonic Imaging

Xinqi Deng,Tingnian Xu,Guoming Huang,Qinghao Li,Liyang Luo,Yue Zhao,Zhengjie Wu,Jun Ou-Yang,Xiaofei Yang,Mingxing Xie,Benpeng Zhu
DOI: https://doi.org/10.1109/tuffc.2020.3028505
2021-04-01
Abstract:Recently, super-harmonic ultrasound imaging technology has caused much attention due to its capability of distinguishing microvessels from the tissues surrounding them. However, the fabrication of a dual-frequency confocal transducer is still a challenge. In this work, 270- $mu text{m}$ PMN-PT single crystal 1–3 composite and 28- $mu text{m}$ PVDF thick film, acting as transmission layer and receiving layer, respectively, are integrated in a novel co-focusing structure. To realize delicate wave propagation control, microwave transmission line theory is introduced to design such structure. Two acoustic filter layers, 13- $mu text{m}$ copper layer and 39- $mu text{m}$ Epoxy 301 layer, are indispensable and should be added between two piezoelectric layers. Therefore, an acoustic issue can be overcome via an electrical method and the successful achievement of a dual-frequency (5 MHz/30 MHz) ultrasound transducer with a confocal distance of 8 mm can be realized. The super-harmonic ultrasound imaging experiment is conducted using this kind of device. The 3-D image of 110- $mu text{m}$ -diameter phantom tube injected with microbubbles can be obtained. These promising results demonstrate that this novel dual-frequency (5 MHz/30 MHz) confocal ultrasound transducer is potentially usable for microvascular medical imaging application in the future.
What problem does this paper attempt to address?