Optical Scanning Analysis of Static Samples by Compact Laser Induced Breakdown Spectroscopy
Yaxiong He,Yong Zhang,Chuan Ke,Qifan Wen,Shu Liu,Dongye Zhao,Min Xu,Tao Xu,Yong Zhao
DOI: https://doi.org/10.1109/tim.2023.3300425
IF: 5.6
2023-08-16
IEEE Transactions on Instrumentation and Measurement
Abstract:In the field of quantitative analysis utilizing laser induced breakdown spectroscopy (LIBS), ensuring the stability and precision of the technique is a crucial consideration. This is particularly pertinent for certain engineering applications, where utilizing electric translation stages for sample scanning is not feasible. In this study, a compact optical scanning LIBS analysis system was developed using a deflection mirror design to replace the translation platform. The system achieved rapid scanning and precise analysis, with a total weight of less than 5 kg. Moreover, to quantitatively analyze Mn, Ni, Cr, Cu, and other elements in various alloy steels, a general method for selecting spectral analysis line combination was proposed. The calibration model was established utilizing 16 alloy steel samples, and its validity was assessed through a verification set of seven samples (with element content ranging from 0.021 to 2.0 wt%). The results indicate a high degree of accuracy. The calibration model determination coefficients ( ) for Mn, Ni, Cr, and Cu elements exceeded 0.99. The root-mean-square error of validation (RMSEV) was less than 0.05, and the relative standard deviation (RSD) was not more than 4%. Additionally, the relative prediction errors for Mn range from 2.98% to 67.44%, for Ni range from 0.57% to 15.93%, for Cr range from 3.91% to 55.56%, and for Cu range from 0.95% to 38.1%.
engineering, electrical & electronic,instruments & instrumentation