Measuring Method and Error Aoalysis of the Caloricity of Natural Gas with Water-Flow Calorimeter

Hongbing Chen
2008-01-01
Abstract:An introduction to the measuring method and the measuring principle of the caloricity of natural gas with water-flow calorimeter is presented,and an analysis is made on the errors being tested.The influence made by the caloricity of natural gas is compared and they are very different.The sources of the measuring errors in the testing is analyzed from the condensate water,the temperature difference,the gas capacity,gas coefficient,etc.The measurement error of the condensate water is one of the most important factors.Corresponding methods to reduce measurement errors are put forward,including extending measurement cycle,decreasing measurement error of the condensate water and the temperature difference.Experiment results show that these methods can effectively reduce measurement error.