Experimental Analysis of Conductance Correlation Flowmeter in Gas-water Two-phase Flow

HUANG Chunhui,HU Jinhai,LIU Xingbin,ZHANG Puwang,SHAN Fujun,WANG Yanjun
DOI: https://doi.org/10.3969/j.issn.1004-1338.2011.06.003
2011-01-01
Abstract:The experiment of conductance correlation flowmeter is carried out on the simulation multi-phase flow experiment facility under the condition of gas-water two-phase flow.The response regulations of the flowmeter in the gas-water two-phase flow are summarized.The analyses of experiment data show that the flowmenter's working range in gas-water two-phase flow is the gas flowrate below 15 m3/d.The flowmeter has different response regulations at low flowrate and high flowrate.When the total flowrate of gas-water two-phase flow is above 20 m3/d and gas flowrate is below 15 m3/d,the correlation flowrate increases linearly with the increases of total gas-water flowrate.When the total flowrate is below 20 m3/d,the correlation flowrate is more sensitive to gas flowrate,and is largely affected by gas slippage.The lower the flowrate is,the more obvious the gas slippage is.When the total flowrate is above 20 m3/d,the tool coefficient in gas-water two-phase flow is almost equal to the tool coefficient in oil-water two-phase flow.The correlation velocity in gas-water two-phase flow is obviously higher than the correlation velocity in oil-water two-phase flow with the same flowrate.