The dynamic response of basket turbine flowmeters in a gas-water two-phase flow

Liu Xingbin,Ping Lin,Huang Chunhui,Song Chungao
DOI: https://doi.org/10.3969/j.issn.1004-9134.2010.04.022
2010-01-01
Abstract:The experiment of turbine flowmeter is carried out with the simulation multi-phase flow experiment facility under the condition of gas-water two phase flow.The response regulations of turbine flowmeter in the gas-water two-phase flow are summarized.The turbine flowmeter has different response regulations at low flowrate and high flowrate.When the water flowrate is above 10 m3/d and the gas flowrate is constant,the response of turbine flowrate is liner to the water flowrate.Further more,along with the increasing of gas flowrate,the K value of turbine flowmeter increases.Under the low fluid volume(<10 m3/d),the response of turbine flowrate is more sensitive to gas flowrate.When gas flowrate is above 20 m3/d,the turbine flowmeter is not sensitive to water flowrate.When the water flowrate is lower than 10 m3/d,the fluctuation of turbine flowmeter response is large;when the water flowrate is above 10 m3/d,the fluctuation of turbine flowmeter response decreases along with the flowrate increasing.