Research of setup errors for left breast cancer radiotherapy in deep inspiration breath hold based on CBCT and OSMS

李克,李勇,邱凌平,张战民,黄焱秋,阙与清
2022-01-01
Abstract:Objective Based on CBCT and OSMS image registration, the setup accuracy between two kinds of fixation methods in deep inspiration breath hold (DIBH) intensity-modulated radiotherapy for left breast cancer were compared. Methods A total of50 patients with left breast cancer treated in our hospital from October 2019 to December 2021 were selected and divided into two groups according to the fixation methods: Klarity multifunctional positional fixation plate (Group A) and Klarity multifunctional positional fixation plate combined with head shaping pads (Group B). Both groups were treated with DIBH mode for whole breast radiotherapy. The registration errors of the two groups in six dimensions were obtained through the rigid registration of CBCT and positioning CT images. The whole treatment process was monitored online by an Optical surface monitoring system (OSMS). Results Group A and Group B took the absolute values of the registration errors in the six dimensions of Lat, Lng, Vrt, Pitch, Roll and Rtn were as follows: (1.45±1.15) mm and (0.93±0.68) mm, (2.21±1.55) mm and (1.39±1.08) mm, (1.68±1.38) mm and (1.45±1.05) mm, (0.67±0.59)° and (0.48±0.44)°, (0.86±0.70)° and (0.70±0.59)°, (0.33±0.67)° and (0.12±0.38)°; There were statistically significant in the directions of Lat, Lng, Pitch, Roll and Rtn (P < 0.05). And there was no statistically significant in the direction of Vrt (P > 0.05). Conclusion Multifunctional positional fixation plate combined with head shaping pad fixation technology can reduce the setup error, and have good clinical application value. It is recommended to use DIBH technology for left breast irradiation.
What problem does this paper attempt to address?