A Low Noise TIA with Continuous Time-Gain Compensation for Ultrasound Transducers

Xinwei Yu,Fan Ye,Junyan Ren
DOI: https://doi.org/10.1109/APCCAS55924.2022.10090351
2022-01-01
Abstract:This paper presents a low noise power-efficient TIA for capacitive-micromachined-ultrasonic transducers (CMUT) to ensure the high-quality ocular ultrasound images. Implemented in CMOS technology, the TIA adopts a source input structure to achieve broadband impedance matching. The use of noise cancellation and capacitive feedback optimizes the noise performance of the amplifier. A continuous time-gain compensation circuit is built in the TIA by adjusting the positive and negative resistance loads, which are composed of the voltage -controlled MOS array. Simulation results show that the TIA achieves a <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\boldsymbol{1.9}\mathbf{pA}/\sqrt{}$</tex> Hz input referred noise at 18MHz center frequency of the echoes while consuming only 0.18mW.