A Versatile Biaxial Stretching Device for in Situ Synchrotron Radiation Small- and Wide-Angle X-Ray Scattering Measurements of Polymer Films.

Jianhe Zhu,Shenghui Liu,Yimin Lu,Hong Cheng,Xueqing Han,Liangbao Liu,Lingpu Meng,Wancheng Yu,Kunpeng Cui,Liangbin Li
DOI: https://doi.org/10.1063/5.0130284
IF: 1.6
2023-01-01
Review of Scientific Instruments
Abstract:A biaxial stretching device is designed and developed for the real-time structural measurements of polymer films. This device adopts a vertical layout to perform real-time x-ray scattering measurements. It has a maximum stretching ratio of 8 × 8 in two perpendicular directions. Its maximum experimental temperature and stretching rate are 250 °C and 100 mm/s, respectively. The control accuracies of the experimental temperature and stretching rate are ±1 °C and 0.01 mm, respectively. All the parameters related to film biaxial processing, such as stretching speed, stretching ratio, and temperature, can be independently set. The device feasibility is demonstrated via a real-time experiment in a synchrotron radiation beamline. Wide-angle x-ray diffraction, small-angle x-ray scattering, and stress–strain data can be simultaneously obtained during various stretching modes. The proposed device fills the gap between the synchrotron radiation x-ray scattering technique and the biaxial stretching processing of polymer films. This device will play an important role in improving the understanding of the physics behind biaxial polymer processing.
What problem does this paper attempt to address?