Design of a Measuring Device and Experimental Study into the Relationship between Temperature and the Density of Alkali Metal-Vapor

Huining Shang,Sheng Zou,Wei Quan,Binquan Zhou,Shun Li,Weiyong Zhou,Fengwen Zhao
DOI: https://doi.org/10.3390/photonics10020112
IF: 2.536
2023-01-01
Photonics
Abstract:The temperature of the alkali metal cell, which affects the density of the alkali-metal vapor and the gas pressure, is usually difficult to measure directly. However, the temperature of the alkali-metal cell and the density of the alkali-metal vapor are important parameters that affect the performance of the atomic sensor. In this paper, a device that can directly measure the internal temperature of an alkali metal cell in real time is designed for the first time to explore the relationship between alkali-metal vapor density and temperature. Alkali-metal vapor density is measured using the absorption spectrum. The pressure broadening model, combined with the transition of four hyperfine levels, was used to fit the absorption line of Rb-87 D1 under the action of 700 Torr N-2, and a good fitting effect was obtained. The experimental results show that the density of Rb-87 is less than the value calculated by the empirical formula. Based on the experimental results, we give the calculation formula of Rb-87 density with an uncertainty of only 4% and obtain the temperature dependence index of the line width and linear displacement of Rb-87 in N-2 by analyzing the absorption spectrum.
What problem does this paper attempt to address?