A Sensitivity Dependence Study for the Chip-Scale Atomic Magnetometer on the Atomic Vapor Cell Size

Hao WANG,Bing XIONG,Zhi-biao HAO,Yi LUO,Jing QI,Chang-zheng SUN,Lai WANG,Jian WANG,Yan-jun HAN,Hong-tao LI
DOI: https://doi.org/10.13385/j.cnki.vacuum.2015.04.16
2015-01-01
Abstract:The shot-noise-limited sensitivity variation due to the cells size was analyzed with the spin relaxation time of alkali atoms and the theoretical relationship was set up. According to the relationship, we calculated the optimum Ar pressure of differ-ent 87Rb cells sizes at 383.15 K. The optimum Ar pressure increases rapidly with the cells size decreases. The magnetic sensitivi-ty limit was also calculated. The results show that the sensitivity drops rapidly with the cell size reduction. When the diameter of the cell decreases from 1 to 0.1 cm, the sensitivity drops from 0.4 to 15 pT Hz-1/2.