Case Study of Electromagnetic Interference in Surgery Environment

Zhao Chen,Johan Catrysse,Ronny Deseine,Tim Claeys,Davy Pissoort
DOI: https://doi.org/10.1109/et55967.2022.9920274
2022-01-01
Abstract:In this paper, a case of electromagnetic interference (EMI) in medical display systems is reported. It concerns the nearby use of an electrosurgical knife disturbing video-monitors, interconnected by Ready-Made Connecting Devices (RMCD), in a surgery room. Based on full wave simulations, three parameters of these RMCD in a medical display installation, including the total length, location of connecting and the number of GND reference - wires, are identified. The simulation results demonstrate that these parameters can not only lead to EMI but are also influencing the characteristics of such EMI.
What problem does this paper attempt to address?