Application of transition-edge sensor for micro-X-ray fluorescence measurements and micro-X-ray absorption near edge structure spectroscopy: a case study of uranium speciation in biotite obtained from uranium mine

Takumi Yomogida,Tadashi Hashimoto,Takuma Okumura,Shinya Yamada,Hideyuki Tatsuno,Hirofumi Noda,Ryota Hayakawa,Sayuri Takatori,Shinji Okada,Tadaaki Isobe,Takahiro Hiraki,Toshiki Sato,Yuichi Toyama,Yuto Ichinohe,Oki Sekizawa,Kiyofumi Nitta,Yuichi Kurihara,Shigeru Fukushima,Tomoya Uruga,Yoshihiro Kitatsuji,Yoshio Takahashi
DOI: https://doi.org/10.1039/d4an00059e
2024-03-12
The Analyst
Abstract:In this study, we successfully applied a transition-edge sensor (TES) spectrometer as a detector for microbeam X-ray measurements from a synchrotron X-ray light source in hard X-ray region to determine uranium (U) distribution at the micro-scale and its chemical species in biotite obtained from the U mine. It is difficult to separate the fluorescent X-ray of the U Lα1 line at 13.615 keV from that of the Rb Kα line at 13.395 keV in the X-ray fluorescence spectrum with an energy resolution of approximately 220 eV of the conven-tional silicon drift detector (SDD). Meanwhile, the fluorescent X-rays of U Lα1 and Rb Kα were fully separated by TES with 50 eV energy resolution at the energy of around 13 keV. The successful peak separation by TES led to an accurate mapping analysis of trace U in micro-X-ray fluorescence measurements and a decrease in the signal-to-background ratio in micro-X-ray absorption near edge structure spectroscopy. Thus, it could be a powerful tool for studying the U distribution and speciation in various environmental samples.