Simulation of the ion trajectories in the ion detection analysis

WANG Yue-Jie,FENG Ming-Chun,DONG Ke-Xiu,SHI Yong-Hua,SHEN Cheng-Yin,CHU Yan-Nan
DOI: https://doi.org/10.3969/j.issn.1000-0364.2018.02.029
2018-01-01
Journal of Atomic and Molecular Physics
Abstract:A conventional and latest Monte Carlo method for simulating the motions of ions in the uniform electro-static fields has been developed and implemented in a user program for Simon 7.0.According to the new ion-molecular collision model and algorithm,the user's collision algorithm program was used to simulate the H3O+motions in the uniform electrostatic field at 1.0 Torr and 298.15 K in the drift tube of Proton Transfer Reaction Mass Spectrometry,which is an example of the ion detection analysis.Based on the ion trajectories simulation and the statistical approaches,the theoretical basis and reference for the optimization of the ion detection analysis to enhance the rate of ion transmission was provided.It is very promising to become a appropriate tool to improve the sensitivity of the ion detection analysis or the similar detecting instruments in the near future.
What problem does this paper attempt to address?