Simulation Study On Linear Ion Trap With Asymmetric Triangular Electrodes
Chen Xin,Gu Lian-Cheng,Xu Fu-Xing,Fang Xiang,Chu Yan-Qiu,Xiao Yu,Zhou Ming-Fei,Ding Chuan-Fan
DOI: https://doi.org/10.19756/j.issn.0253-3820.191314
IF: 1.193
2019-01-01
Chinese Journal of Analytical Chemistry
Abstract:The triangular-electrode linear ion trap (TeLIT) has many advantages such as simple structure , easy processing and high quality analysis performance. In this work , the internal electric field composition and mass analysis ability of two asymmetric TeLIT were explored by using theoretical simulation method , including (1) two electrodes in the direction of the ion exiting had the same angle but differ with the other two identical electrodes, and (2) one electrode in the direction of the ion exiting had the different angle compared with the other three identical electrodes. The simulation results showed that when one triangular-electrode with the angle of 137 degrees in the direction of the ion exiting and the other three with the angle of 140 degrees , odd and even order fields were introduced simultaneously in the linear ion trap , and when the scanning speed were 1900 Da/s and at m/z 609 Da, the mass resolution could reach 7186. The sensitivity and accuracy of the linear ion trap was also improved.