Abstract:XFCT (X-ray fluorescence CT) is a new type of medical imaging system, which can present molecular and functional information in organisms. XF (X-ray fluorescence) imaging has the advantages of low cost, easy storage and transportation of contrast agents and low internal radiation. However, the mechanical collimation limits the detection efficiency of the system. Compton camera is an imaging system composed of double-layer detectors, which can obtain the direction of incident rays without a collimator. So combining the advantages of Compton camera system and XF photon imaging is a solution worth exploring. Timepix3 is one of the most advanced high resolution photon counting detectors, which has well time resolution and can obtain the depth position where particle deposit energy. In this work, we use Si-Timepix3 single-layer detector to build up a Compton camera system, in order to detect the 43keV XF photons from Gd solution excited by X-ray. We establish the first real experimental platform to verify the feasible of XFCC (X-ray Fluorescence Compton Camera) system and obtain the reconstruction images.
What problem does this paper attempt to address?