Characterization of 1mm Cross-strip 3D CZT Detectors for PET Imaging Application

Yingguo Li,Changxu Pei,Meilou Liu,Jing Wen,Qinghua Zhang,Chuan Huang,Yongzhi Yin,Gongping Li,Ximeng Chen
DOI: https://doi.org/10.1109/nss/mic44867.2021.9875495
2021-01-01
Abstract:This paper studies a cross-strip cadmium zinc telluride (CZT) PET detector with a size of 18mm×18mm×5mm, which can provide 3-D position of 511 keV gamma-photons. We used charge-sensitive preamplifiers and filtered amplification modules to process the 32 electrodes, then sampled the data with high-speed analog-to-digital converters, afterwards the signal was processed online using the ADC built-in FPGA. Through experimental tests, we have obtained the complete signal pulse waveform of the CZT detector, its peak value and time information. The CZT detector exhibits an energy resolution of 3.92% for the energy spectrum of 511keV. The 3D position information of the gamma-ray interaction is obtained, and it can be about the sub-millimeter.