A Novel Approach: Successful Management of Vasectomy Reversal with a Three-Dimensional Digital Image Microscope System
Peng Li,Erlei Zhi,Chencheng Yao,Shujie Xia,Zheng Li
DOI: https://doi.org/10.4103/aja.aja_15_20
2021-01-01
Abstract:to perform a vasovasostomy with the 3D digital image microscope in humans. In August 2018, a 44-year-old male with an 18-year-old son from his first marriage was admitted to our hospital for obstructive azoospermia with a history of bilateral vasectomy 18 years ago. He remarried 4 years ago, and his wife had no infertility factors. Physical examination showed bilateral palpable vasal defects, no sperm granuloma, and normal testes. Semen analysis indicated that no sperm was detected, and scrotal ultrasound evaluation demonstrated dilated vas deferens and epididymal tubules. The sex hormones were all in the normal range. We spoke with the patient and his wife regarding the alternative procedure of assisted reproductive technology that may be helpful for the treatment of his infertility as well as the feasibility and safety of microsurgical vasovasostomy; eventually, they agreed to receive this procedure using the 3D digital image microscope. The procedure was approved by the Ethics Committee of Shanghai General Hospital (Shanghai, China), and informed consent was obtained from the patient. The procedure was performed under the 3D digital image microscope (Mitaka Kohki, Mitaka, Japan) with ×1.9–39.3 magnification, and the procedure is described in detail in the following steps: (1) the 3D digital image microscope was placed over the surgical field, the working distance was 350–1000 mm, and two surgeons stood on each side of the operating table in a natural position and viewed opposite monitors on both sides (Figure 1); (2) under the microscope with ×2–4 magnification, a 3-cm vertical middle scrotal incision was made, and the testes were delivered; (3) the vas deferens was exposed at the prior vasectomy site, and both ends of the vas deferens were subsequently transected; (4) the abdominal end of the vas deferens was cannulated with a 24-gauge angiocatheter sheath to confirm the patency by injecting diluted methylene blue into the distal part of the vas deferens; (5) the fluid from the testicular end of the vas deferens was examined under light microscopy intraoperatively, and a few motile sperm were found; (6) both vasal ends were fixed with the microspike approximator clamp (ASSI Corporation, Westbury, NY, USA), and six microdots were marked on each side5 (Supplementary Figure 1a); (7) under the microscope with ×30–39.3 magnification, three interrupted sutures of 10-0 single-armed monofilament proline were placed through only the mucosal layer of the vas deferens, and then the mucosal sutures were tied with a microsurgeon’s knot (Supplementary Figure 1b and 1c); Dear Editor, Microsurgical vasovasostomy, most commonly performed for vasectomy reversal, remains the most successful procedure for restoring patency to the vas deferens with the return of sperm to the ejaculate. Although the excellent results depend on the surgeon’s skill and technique, optimal vision and ergonomics are also crucial factors contributing to the achievement of good results during male infertility microsurgery. Recently, the use of a new revolutionary technology, called three-dimensional (3D) digital image microscope with a video-assisted telescope operating monitor, may fill the gap between the need for good surgical vision and maintenance of a comfortable posture during surgery and may improve surgeon comfort. We present the first case report of vasovasostomy performed using the 3D digital image microscope system and discuss advantages and disadvantages compared with the use of an operating microscope. The 3D digital image microscope system, which consists of a scope located over the surgical field that is supported by a mobile mechanical arm and projects to a screen, can easily fit all kinds of observation angles and can produce high-quality 3D videos of a small surgical field, with a high magnification capacity (×1.9 to ×39.3) and a long working distance from 350 to 1000 mm; this system allows surgeons to work in a setting similar to that of endoscopic surgeries, with possibilities of application similar to those of operating microscopes. The application of an ordinary video camera has been reported for microvascular anastomosis;1 furthermore, the 3D video-assisted telescope operating system has been used in cranial surgery and spinal surgeries.2–4 Barros et al.1 performed cranial surgery using a video-assisted telescope operating monitor and demonstrated that this system might have many advantages regarding ergonomics, versatility, and field depth compared with an operating microscope.1 We first performed vasovasostomy on rats to validate the feasibility and safety of this system. The mechanic patency rate was 100% (8/8) for 3D digital image microscope-assisted vasovasostomy in the rat model (unpublished data), which revealed that it may be possible LETTER TO THE EDITOR