Abstract:The eddy current testing (ECT) technique is a non-destructive evaluation method that is widely used in a variety of fields for detecting defects and their locations. A particular problem with this method is that variations in probe lift-off affect the ECT results. This is a well-known problem during online defect inspection and quantification, for example in train rail inspections. To this end, this study proposes a novel lift-off invariance transformation method for reducing the undesired noise caused by lift-off The efficacy of the proposed method has been compared with the dual-frequency excitation method and the validity of the transformed signals has been verified by a novel model-free method: support vector regression optimised by particle swarm optimisation (PSO-SVR). Experimental results prove the better performance of the proposed method in transforming the original signals to zero lift-off equivalent signals, which could play a significant role in subsequent quantitative defect estimation.
What problem does this paper attempt to address?