A Method To Predict Fatigue Performance Of Artificial Blood Vessels

Zhao Hui-Jing,Wang Lu,Li Yu-Ling,Liu Xiao-Yan,King Martin
2008-01-01
Abstract:Artificial blood vessels include three parts as vascular prostheses, endovascular prostheses, as well as tissue engineering grafts. The dynamic fatigue performance of artificial blood vessels can be predicted by an accelerated fatigue tester. The EnduraTec fatigue tester can provide pulsatile pressure to the scaffold samples. The systolic and diastolic pressure can be selected as 200 and 140mmHg respectively. The pressure wave form is sine wave. The frequency of the tester can be adjusted from 0 to 200Hz. The temperature of the fluid inside the Tester was maintained as 37 degrees C. Twelve samples can be tested on the tester simultaneously. To characterize fatigue performance of scaffold, the changes of tubular diameter, fabric count, and water permeability before and after fatigue test can be used. In this study, several, prototype scaffold samples of stent-graft were used to validate the method for predicting fatigue performance as well as the equipment. It was found that samples experienced fatigue during the test by the EndurTec fatigue tester. The parameters selected to characterize fatigue performance can well indicate fatigue process of the scaffolds. In conclusions, this equipment and the method is reliable and effective for predicting the fatigue performance of textile scaffold of the stent-graft, they are also suggested to be used to evaluate fatigue behaviors of tissue engineering scaffold and grafts.
What problem does this paper attempt to address?