In vitro Evaluation of the Microstructural Vascular Scaffold with Micro-Tip Direct Writing Technology

sun pingping,hu chuan,zhou jianye,wu rendong,shen ya,hu shengshou
DOI: https://doi.org/10.3969/j.issn.1002-3208.2011.02.01.
2011-01-01
Abstract:Objective To evaluate the mechanical properties and biocompatibilities of the vascular scaffold manufactured with micro-tip direct writing technology.Methods Mechanical properties of the vascular scaffold were measured by the parameters of radial compliance,suture strength and bursting pressure separately.The blood compatibilities of the scaffold were analyzed by hemolytic rate test,in vitro dynamic blood clotting test,ant platelet adhesion experiment.Cell compatibility was evaluated by MTT measurement and morphological observation.Results The scaffold showed radial compliance of(4.03±0.56)%/100mmHg,burst strength of(102±8)kPa,and suture retention strength of(204.5±72.1)N/cm2.The hemolytic rate was 1.75%,being lower than 5%,which was ruled by ISO.The anti-coagulant property of the scaffolds was significantly better than that of glass slides in dynamic coagulation tests(P<0.05).The anti-coagulant property of the scaffold was also confirmed by less platelet adhesion.MTT assay results showed that the cytotoxic grade of the novel polyurethane was from 0 to 1.Cell morphology observation showed that the L929 cells were spindle-shaped or triangular with good stretch.Conclusion The scaffold manufactured with the new technique processes good mechanical properties,blood compatibility and cell compatibility.