An Experimental Modal Investigation of the Dosimetry Parameters for Self-Expandable Esophageal Stent Loaded with I-125 Seeds
Lei Lin,Junjie Wang,Ruijie Yang,Yuliang Jiang,Weijuan Jiang,Shukun Jiang,Ziyi Liu,Min Wang
DOI: https://doi.org/10.1016/j.brachy.2015.02.311
IF: 2.441
2015-01-01
Brachytherapy
Abstract:With the widespread adoption of self-expandable esophageal stent loaded with 125I seeds, a number of parameters of the irradiation stent dosimetry were needed to identify. The aim of this study was to investigate potential dosimetry distribution for self-expandable esophageal stent loaded with 125I seeds and to propose standard dosimetry parameters. Six models of stents (Nanjing Micro-Tech Co Ltd, Nanjing, China) which were 8.0cm length×2.0cm diameter with 1.5cm spacing of the seed center (A), 8.0cm length×2.0cm diameter with 1.0cm spacing of the seed center (B), 8.0cm length×1.3cm diameter with 1.0cm spacing of the seed center (C), 8.0cm length×2.4cm diameter with 1.0cm spacing of the seed center (D), 12.0cm length×2.0cm diameter with 1.0cm spacing of the seed center (E), and 16.0cm length×2.0cm diameter with 1.0cm spacing of the seed center (F), were studied. The treatment planning system (TPS) used to calculate the dose delivery at specific points. The points were defined at polar angles of 0-90°in 10°increments and at radial distances from outer surface of the stent were 0.5, 0.75, 1, 1.25, 1.5, 2, 2.5, 3, 3.5, 4, 5, and 7cm. A cross section of the detector was arranged. The thermoluminescence detector (TLD) used for this study was LiF:Mg,Cu,P (GR 200A). The detectors were defined at polar angles of 0-90°in 10°increments and at radial distances from outer surface of the stent were 0.5, 0.75, 1, 1.25, 1.5, 2, 2.5, 3, 3.5, 4, 5, and 7cm. A cross section of the detector was arranged. The readout system was a TL Reader (Thermo, USA). The heating cycle consisted of preheat, readout, anneal and cool segments. The Monte Carlo (MC) were used for the dosimetry simulations around the irradiation stent. The range for photon and electron transport extends from 1 KeV to 100 MeV. The irradiation stents have been simulated in the centre of a spherical phantom of water with 20 cm radius, large enough to consider all the scattering effects of the surrounding medium. The detectors were defined at polar angles of 0-90°in 10°increments and at radial distances from outer surface of the stent were 0.5, 0.75, 1, 1.25, 1.5, 2, 2.5, 3, 3.5, 4, 5, and 7cm. A cross section of the detector was arranged. Compare all the results of these three methods and propose standard dosimetry algorithm. Source strength lineal density is equal for peripheral sources to interior sources. The results show the schematic diagram of the irradiation stents geometry and algorithm used with algorithm for dose calculations to a point approximation. The superposition principle assumed the total dose distribution to be a summation to the dose contributions of all individual sources loaded a stent. When the seed activity was ≤0.6mCi, the isodoses were not well-distrubuted in each plane. At the point of 145Gy from the stent outer surface, the isodose was discontinuous which leads to radiology cold on the surface of the stent. When the seed activity was ≥0.7mCi, the isodoses well-distrubuted and trended to nearly circular shaped in each plane, especially when the seed activity was ≥0.9mCi. At the point of 145Gy from the stent outer surface, the isodose was continuous. The “system” described allowed the users to administer a safe dose easily to the patient for the treatment of the tumor by providing the rules for arrangement and placement of the sources loaded with the stent. When the activity of 125I seeds were ≥0.7mCi, the isodoses of 145Gy were continuous whthout dosimetric cold spots where are the location of the tumor recurrence.