Dosimetry Study of Permanent Radioactive Biliary Stent Loaded with 125I Seeds

Lihong Yao,Lei Lin,Lu Liu,Haitao Sun,Ruijie Yang,Liu Hao,Di He,Junjie Wang
DOI: https://doi.org/10.1016/j.brachy.2016.04.290
IF: 2.441
2016-01-01
Brachytherapy
Abstract:To evaluate the dosimetric characterization of a new designed radioactive biliary stent loaded with 125I seeds and report results in a form suitable for clinical use. Three commonly used stent models were studied. Dose measurements were made using standard methods employing treatment planning system (TPS) calculation, thermoluminscent dosimeters (TLDs) measurement, and Monte Carlo (MC) simulation. A solid paraffin and cylindrical barrel model made of Polymethylmethacrylate (PMMA) were used as a equivalent water phantom for TPS calculation. After a radioactive biliary stent was positioned in the measurement phantom, a CT scan was performed to get the stent images. The images were then transferred to the TPS for treatment planning. The intermediate CT image of the stents was selected to calculate the surface radial dose distribution. Radial dose distribution around the stents was also measured in a special designed PMMA phantom using TLD-GR 200A circular chips. In preliminary experiment, corrections were made to study the TLDs linearity of dose response, energy response, and scaled for exposure duration. MC simulations were performed using the MCNPX v.2.4 general purpose code to calculate the biliary stent three-dimensional dose rate distribution in liquid water. The length of the stents (F = 3189.160, P < 0.001) and the activity of the 125I seeds (F = 811.509, P < 0.001) could both significantly affect the cumulative radial dose distribution of the radioactive stent. For the same length of radioactive stents, the cumulative radial dose of the stent was different (P < 0.05) when the seed activity difference was of 0.1 mCi. The simulations with MCNPX agree in acceptable agreement with the TPS calculations and TLD measurements. The MC simulated uncertainty of the three stent models were all within the scope of the report recommended by AAPM TG-43U1. Depth dose calculations of different length radioactive biliary stents in different activity of 125I seeds were performed at 0.1, 0.5, 1.0, 1.5 and 2.0 cm from surface of the stent. The radioactive biliary stent doses at the reference point could be adjusted by changing the seed numbers, and source activity. The TPS calculation results and the TLD experimentally defined radial dose rate were in good agreement with MCNPX simulations. Determinations of the values of the 3D dose rate in water have been simulated using Monte Carlo simulations.
What problem does this paper attempt to address?