Radiation Dose Distribution of Implanted Metal Stent

XIONG Fei,LI Zhao-bin,JIANG Rui-yao,HUANG Guo-feng,CUI Shao-xiang,FU Shen
DOI: https://doi.org/10.3969/j.issn.1672-4992.2011.01.051
2011-01-01
Journal of Modern Oncology
Abstract:Objective:To determine the scattering effect of implanted metal stent on radiation dose distribution and define the mathematical correction parameters for post-implantation irradiation dose calculation.Methods: The phantom with nickel-titanium alloy stent was scanned by CT-sim with 5mm slice.The CT images were transferred to TPS(treatment planning system) by LANTIS network,and the interest point dose was calculated.According to the scan condition,the radiation was done with 6 MV and 15 MV X-ray.The dose was measured by TLD and FAMAR irradiation chamber.The result calculated from the TPS was compared with the measurement datum.Results: The measurements showed an increase in absorbed dose of up to 3.9% in 6 MV and 6.6% in 15 MV x-rays before the surface of stent,up to 2.8% in 6 MV and 6.3% in 15 MV x-rays after the surface of stent.The errors become smaller with the distance increase.Conclusion: The presence of stent increased the dose,which could lead to radiation damage,so the radiation dose should be amended while making radiotherapy planning.