Design of Mud Density Meter Using Monte Carlo Method

LI Zheng-ping,WU Rui-sheng,YIN Ze-jie,WANG Hu-jian
DOI: https://doi.org/10.3969/j.issn.0253-2778.2005.05.011
2005-01-01
JUSTC
Abstract:The design and implementation of a γ-ray mud density meter is presented.The influence of various factors on measurement precision,sensitivity and radiation pollution was simulated and analyzed by using Geant4 software package.On the basis of simulation,the physical design of the density meter is presented,including the selection of radiation source,the calculation of source intensity and design of lead container as well as the design of system hardware and software.Experimental results show that the measurement precision is better than 1%,the range of measurement is from 1.00g/cm~(3) to 1.60g/cm~(3),and the radiation pollution meets the national criterion.
What problem does this paper attempt to address?