Design of a Measurement Unit for Measuring the Neutron, Gamma Dose Equivalent Meanwhile Based on Single-sphere Multi-detector Neutron Spectrometer and GM Counter

Ming-ming YANG,Jia-tao SONG,Jian-wei LI,De-yuan LI,Bo-yuan WANG,Jian-zhong LIU
DOI: https://doi.org/10.3969/j.issn.0258-0934.2017.02.020
2017-01-01
Abstract:This article describes a unit used for measuring the dose equivalent of neutron and gamma ray of the mixed radiation field at the same time,which is composed of single -sphere multi-detector neutron spectrome-ter and GM counting tube.The detector of single sphere with multi -counter is composed of three He -3 pro-portional counters which is orthogonal.It has been tested in the standard neutron radiation field of Cf -252 and Am-Be,the relative error of the dose equivalent is less than 30%.The measurement of energy response has been tested in the incident photon beam which's energy is between 1 MeV and 14 MeV, the relative error is less than 30%.
What problem does this paper attempt to address?