Cosmic-muon Intensity Measurement and Overburden Estimation in a Building at Surface Level and in an Underground Facility Using Two BC408 Scintillation Detectors Coincidence Counting System

Weihua Zhang,Kurt Ungar,Chuanlei Liu,Maverick Mailhot
DOI: https://doi.org/10.1016/j.jenvrad.2016.06.011
IF: 2.655
2016-01-01
Journal of Environmental Radioactivity
Abstract:A series of measurements have been recently conducted to determine the cosmic-muon intensities.and attenuation factors at various indoor and underground locations for a gamma spectrometer. For this purpose, a digital coincidence spectrometer was developed by using two BC408 plastic scintillation detectors and an XIA LLC Digital Gamma Finder (DGF)/Pixie-4 software and card package. The results indicate that the overburden in the building at surface level absorbs a large part of cosmic ray protons while attenuating the cosmic-muon intensity by 20-50%. The underground facility has the largest overburden of 39 m water equivalent, where the cosmic-muon intensity is reduced by a factor of 6. The study provides a cosmic-muon intensity measurement and overburden assessment, which are important parameters for analysing the background of an HPGe counting system, or for comparing the background of similar systems. (C) 2016 Elsevier Ltd. All rights reserved.