Development of a low background liquid scintillation counter for a shallow underground laboratory

J. L. Erchinger,C. E. Aalseth,B. E. Bernacki,M. Douglas,E. S. Fuller,M. E. Keillor,S. M. Morley,C. A. Mullen,J. L. Orrell,M. E. Panisko,G. A. Warren,R. O. Williams,M. E. Wright
DOI: https://doi.org/10.1016/j.apradiso.2015.08.027
2015-12-21
Abstract:Pacific Northwest National Laboratory has recently opened a shallow underground laboratory intended for measurement of low-concentration levels of radioactive isotopes in samples collected from the environment. The development of a low-background liquid scintillation counter is currently underway to further augment the measurement capabilities within this underground laboratory. Liquid scintillation counting is especially useful for measuring charged particle (e.g., $\beta$, $\alpha$) emitting isotopes with no (or very weak) gamma-ray yields. The combination of high-efficiency detection of charged particle emission in a liquid scintillation cocktail coupled with the low-background environment of an appropriately-designed shield located in a clean underground laboratory provides the opportunity for increased-sensitivity measurements of a range of isotopes. To take advantage of the 35 meters-water-equivalent overburden of the underground laboratory, a series of simulations have evaluated the scintillation counter's shield design requirements to assess the possible background rate achievable. This report presents the design and background evaluation for a shallow underground, low background liquid scintillation counter design for sample measurements.
Instrumentation and Detectors,Nuclear Experiment
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a low - background liquid scintillation counter (LSC) in a shallow underground laboratory in order to improve the sensitivity of measuring low - concentration radioactive isotopes in the environment. Specifically, this research aims to significantly reduce the Minimum Detectable Activity (MDA) by reducing background radiation, so as to be able to more accurately measure the low - level radioactivity of beta - emitters such as tritium (^3H), carbon - 14 (^14C), and strontium isotopes. This improvement is of great significance in fields such as the study of environmental transport mechanisms, bi - radioisotope dating, radioactive fallout monitoring, and nuclear accident emission tracking. ### Key points: 1. **Reduction of background radiation**: - By designing and optimizing the shielding system, the background radiation rate is reduced to 0.01 to 0.07 counts per minute (cpm). - Utilize a 35 - meter water - equivalent overburden, combined with simulation to evaluate the shielding design requirements, to further reduce background radiation. 2. **Reduction of detection limit**: - Use the Currie equation to calculate the Minimum Detectable Activity (MDA): \[ \text{MDA}=\frac{2.71 + 4.65\sqrt{C_bT_b}}{\epsilon\times V_s\times T_s\times60} \] where: - \(C_b\) is the background count rate (cpm) - \(T_b\) is the background count time (in minutes) - \(\epsilon\) is the counting efficiency - \(V_s\) is the sample weight (in grams) - \(T_s\) is the sample count time (in minutes) 3. **Practical application cases**: - **Case 1**: Measuring the tritium background in natural spring water. Using the low - background LSC system, the same detection limit (MDA) as the traditional method can be achieved without electrolytic enrichment, and the total processing time is shortened from about 6 days to less than 1 day. - **Case 2**: Determining strontium - 90 (^90Sr) in food samples. By reducing background radiation, the sample amount can be reduced, thereby simplifying the complex sample pre - treatment process. - **Case 3**: Measuring actinium - 227 (^227Ac) in seawater. Using the low - background LSC system, actinium - 227 can be directly measured without waiting for the growth of daughter products, achieving the same sensitivity as the traditional method. ### Shielding design: 1. **Multi - layer shielding concept**: - Outer lead shielding, inner low - background lead shielding, and high - purity copper shielding. - The copper shielding also forms a light pipe to transmit scintillation light from the sample bottle to the photomultiplier tubes (PMTs). 2. **Neutron shielding**: - Use boric acid polyethylene plates and plastic scintillation panels to reduce the influence of neutrons and muons. 3. **Simulation verification**: - Use GEANT4 software for simulation to evaluate the influence of different shielding material thicknesses on background radiation and optimize the shielding design. Through these designs and optimizations, this research aims to develop a liquid scintillation counter that can operate efficiently and with a low background in a shallow underground laboratory, significantly improving the detection ability of low - level radioactive isotopes.