Characteristics and Application of Elliptical Curved LiF Crystal Analyzer
XIAO Sha-li,SHI Jun,LEI Xiao-ming,QIAN Jia-yu,XU Dong-qiang
DOI: https://doi.org/10.3321/j.issn:1004-924x.2007.06.004
2007-01-01
Optics and Precision Engineering
Abstract:Based on the principle of elliptical focusing,an Xray elliptical curved crystal analyzer whose measurement range,focal length,eccentricity and Bragg angel range are respectively 0.6~6 keV,1350 mm,0.9586 and 30°~60° was designed for investigating the laser producing plasmas.Aimed to the characteristic study of a high density plasma,a experiment was carried out with laser length of 0.35 μm,power of 1.6×1014 W/cm2,and the incident angel on the Ti plane of 45°.The result shows that the sensitivity of elliptical curved LiF crystal has very good detection ability,its calculated wavelength resolution is about 500~1 000 in the range of 0.2~0.35 nm.Experimental results also show that the curved crystal analyzer can be used in spatial resolution measurement for aplanatic property,its sensitivity is higher than a flat crystal.It suggests that curved LiF crystal analyzer is suitable for plasma diagnostic.