Measuring Bone Mineral Density with NMR

LI Cheng-wei,FAN Dong-wei,JIANG Jia-qi,DU Guo-hong,WANG Wei-min
DOI: https://doi.org/10.11938/cjmr20160312
2016-01-01
Abstract:Bone mineral density (BMD) measured by dual energy X-ray absorptiometry is often used as an index for diagnosis of osteoporosis. However, ionization radiation is present in X-rayBMD measurement, and the result can be inaccurate in some caese. In this work, we constructed an NMR equipment that used a Q-switch to achieve echo time as low as 80ms, and thus have the potential to be used forBMD measurement. In theT2 relaxation time spectra of bone measured by this equipment, the peak area at 300~500ms was found to change linearly with theBMD measured by X-ray absorptiometry. Based on this observation, we put forward a method to measurein vitroBMD with NMRT2 relaxation time spectra. This kind of method has advantages such as high measuring speed, no ionization radiation, and high cross-period comparability.