Improvement of Formaldehyde Analysis in Cosmetic Products by a Dual-wavelength Spectroscopy

Ning YAN,Shu-xin ZHANG,Xin-sheng CHAI
DOI: https://doi.org/10.3969/j.issn.1004-4957.2016.11.024
2016-01-01
Abstract:In this paper,a method was presented for the determination of formaldehyde in cosmetics based on a dual-wavelength spectroscopy,in which both absorbances at 414 nm and 500 nm were used.The results showed that the relative standard deviation(RSD) of the method is only 0.46%,while it is 11.5% by the single-wavelength method.The limit of quantitation(LOQ) of the dual-wavelength method is 7.4 mg/kg.The present method is sensitive and accurate,and is suitable for the the determination of low content formaldehyde in cosmetic products.