Development of an Automated Measurement System Based on a Coil Collector and High Performance Liquid Chromatography for Detection of Formaldehyde in Atmosphere

MA Yi-yuan,WANG Hui-xiang,WANG Shao-li,LIU Bin
DOI: https://doi.org/10.3969/j.issn.1003-6504.2010.05.024
IF: 11.4
2010-01-01
Environmental Science & Technology
Abstract:A continuous determination system of atmospheric formaldehyde was established based on high performance liquid chromatography(HPLC). The sampling system employs scrubbing coil to trap atmospheric formaldehyde into an acid solution of 2,4-dinitrophenylhydrazine (DNPH). Reaction between formaldehyde and DNPH produces a stable product which is determined by reverse-phase HPLC and UV detection. All measurement operations are controlled by DasyLab. The automated measurement can be performed at 60min intervals. The detection limit was 0.10μg/L. The concentration of formaldehyde was determined with the method. The average concentration is 4.82μg/L,and showed obvious diurnal variation in October,2007.