Identification of different perfumes based on microchip-based FAIMS technique

WANG Qi,YU Wen-jing,WANG Xiao-zhi,LI Ling-feng,ZHANG Yuan,LI Peng,ZHAO Li-qing,QIAN He
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2013.0177
2013-01-01
Abstract:A method of distinguishing different brands of perfumes by microchip-based high-field asymmetric ion mobility spectrometry(FAIMS) was established.In this method,63Ni was utilized as an ionizing source and pure air as a carrier gas.Both positive and negative ion detection at atmosphere could be achieved at the same time.By optimizing sample introduction temperature at 20 ℃,sample gas at a flow rate of 0.1 L/min and carrier gas at a flow rate of 1.4 L/min,the CV spectra of six brands of perfume were characterized.Unique features for each perfume were extracted for recognition.The relative standard deviation(RSD) of this method was lower than 6%.Compared with other instruments,this method has advantages in its rapid identification of perfumes and other volatile chemicals with hassle-free sample introduction process.Moreover,this work demonstrates the great capability of microchip-based FAIMS technology in the applications for rapid analysis and identification of volatile samples.
What problem does this paper attempt to address?