Based on SIMION Simulation to Identify Peaks of High-Field Asymmetric Waveform Ion Mobility Spectrometry Spectra

Bin LI,Fei TANG,Yue ZENG,Xiaohao WANG
DOI: https://doi.org/10.3969/j.issn.1004-1699.2014.12.005
2014-01-01
Abstract:Due to work in an open atmospheric conditions, High-field asymmetric waveform ion mobility spectrometer (FAIMS)is susceptible to interference by impurities in the air,thus in the actual analysis more than one peak are usually detected,which creates certain difficulties for identification our main target sample ions observed,however,combing mass spectrometer with FAIMS is the most used method at present. To solve structure complex of combination,in this study, through analyzing several major ions of acetone sample which is ionized by UV lamp ionization source,contrast compensa-tion voltage curve with the change of radio frequency voltage amplitude in SIMION simulation and each ion peak of ex-perimental FAIMS spectra respectively,three major ions corresponding to FAIMS spectra preliminary deduces from ac-cording to their parallel relationship:acetone monomer ion[M+H]+,acetone dimer ion[2M+H]+and hydrated protons ( H2 O) n H+. And measurement error of migration zone spacing and effect of radio frequency voltage amplitude,the reason for compensation voltage of simulation and experiment curves approximately parallel is theoretically analyzed. Simulation results and theoretical analysis are consistent,correctness and feasibility of the method is further proofed.
What problem does this paper attempt to address?