Immediate and Short-term Effects of AV-delay Optimization in Patients with Cardiac Resynchronization Therapy by Echocardiography

张军,李金芳,刘丽文,张海滨,朱永胜,王红梅,刘兵
DOI: https://doi.org/10.3969/j.issn.1002-0101.2010.02.010
2010-01-01
Abstract:Objective To observe the effect of optimizing the atrioventricular (AV) delay on the hemodynamics as well as the short-term efficacy in patients with cardiac resynchronization therapy by echocardiography.Methods Twenty-four patients with chronic heart failure who had received cardiac resynchronization therapy (CRT) were divided into optimization group and control group randomly.The optimal AV delay was tested by echocardiography using Ritter's method more than three months after CRT.The parameters of the patients in control group were restored to the initial settings after optimization.Follow-up (FU) was conducted one month later.Results In optimization group the left ventricular filling time (LVFT) significantly increased immediately after optimization and at FU(LVFT:453±86.6ms,490.5±122.4ms vs396.6±126.7ms,P<0.05).The left ventricular end-diastolic diameter (LVEDD) decreased and there was a slightly increase of the LV-ejection fraction(LVEF) at FU (67.4±8.28)mm vs(71.08±8.52)mm,27.58±6.05% vs.25.50±6.99%,P<0.05).In control group the LVFT significantly increased immediately after optimization but not at FU.The LVEDD and LVEF did not change significantly at FU.Conclusions Optimization of AV-delay by echocardiography can improve the systolic and diastolic function and further enhance the efficacy of CRT.