A comparative study of acute hemodynamic effects of interval optimization for cardiac resynchronization therapy patients by echocardiography and QuickOpt function

LIANG Yi-xiu,SU Yan-gang,GONG Xue,PAN Wen-zhi,BAI Jin,WANG Wei,GE Jun-bo
DOI: https://doi.org/10.3969/j.issn.1007-2659.2010.05.007
2010-01-01
Abstract:Objective To investigate the concordance of acute hemodynamic effects of optimization of SAV,PAV and VV delay between echocardiography and QuickOpt function method.Methods A total number of 19 patients implanted with QuickOpt functioned CRT(D) were enrolled.Echocardiography was used to acquire aortic velocity time integral(aVTI)and optimized SAV,PAV,and VV delay.Also QuickOpt function was used to acquire the optimized SAV,PAV,and VV delay and further measurement of corresponding aVTI was made.Results Concordance correlation coefficient(CCC) between the maximum aVTI values acquired by the standard method and those by the IEGM method were 95.95%,97.29%,92.45% respectively,by data analysis of the SAV,PAV and VV delays(P<0.01).The optimal SAV,PAV and VV interval acquired by echocardiography method were not different from those by QuickOpt method (all P>0.05).Optimization of interval with QuickOpt method saved a lot of time compared with echocardiography method(1.28±0.45 min vs 43.60±11.07 min,P<0.01). Conclusion QuickOpt function is an alternative method for the optimization of AV,PV and VV delay settings in patients with CRT(D) which provides similar effect on acute hemodynamic analysis and saves a lot of time compared with standard echocardiography method.
What problem does this paper attempt to address?