Value of Amplatzer Occlusion Device of Atrial Septal Defect Guided by Echocardiography

张军,李军,田雪,朱永胜,姚志勇,朱霆,张玉顺,李寰,代政学,李媛,王晓燕,刘建平
DOI: https://doi.org/10.3969/j.issn.1002-0101.2001.06.007
2001-01-01
Abstract:Objective:To probe into the guiding meaning of echocardiography for occlusion of atrial septal defect(ASD)with different size and edge by use of Amplatzer septal occluder.Methods:Fifty three patients with ASD were occluded by use of Amplatzer Septal occluder after transesophageal and transthoracic echocardiographic detection or measurement of expanding size of ASD with balloon filling.Results:Twenty cases of ASD(16.8±8.8mm)with solid edge were occluded with selected Amplatzer septal occluder depending on echocardiographic examination.Thirty one patients with soft and flappy edge ASD(22.4±6.9mm)were closed by use of the Amplatzer septal occluder,which sizes were decided according to ASD expanding size by balloon.Correlation between the echo size of ASD and Occluder size,real measurement of balloon and balloon size measured in X ray were fine,r=0.930(P<0.01),r=0.921(P<0.001)and r=0.876(P<0.001)respectively.After eliminating ASD with soft and flappy edge,the correlation became excellent,r=0.984(P<0.001),r=0.959(P<0.001)and r=0.948(P<0.001)respectively.Conclusions:Echocardiography can be used for guidance of Amplatzer occlusion device of ASD with solid edge without the measurement of balloon expanding size of ASD.However,the expanding size should be measured in ASD with soft and flappy edge.