Changes of pseudophakic accommodation of aspheric intraocular lens designs in phacoemulsification

Yuqin Zhao,Yingjun Li,Longshan Jin
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2015.02.007
2015-01-01
Abstract:Objective To compare the pseudophakic accommodation of different designs of aspheric intraocular lens (IOL) in phacoemulsification.Methods One hundred and twenty-six eyes with cataract surgery were randomized into 2 groups:an Acrysof IQ SN60WF (Alcon Laboratories,Fort Worth,Texas,USA) IOL group (59 eyes) and Akreos Adapt AO (Bausch & Lomb.,Rochester,NY,USA)IOL group (67 eyes).Patients were examined using anterior segment optical coherence tomography (AS-OCT),including anterior chamber depth (ACD),pupil diameter,and amplitude of pseudophakic accommodation,preoperatively and 6 months postoperatively.Results The IQ IOLs and AO IOLs group of mean pseudophakic accommodation was (1.26±0.34)D and (1.09±0.51)D,respectively,there was statistically significantly different (t =2.314,P =0.041).The mean pupil diameter measurements for IQ group and AO IOL group were (0.52±0.20)mm and (0.49±0.16)mm,respectively,which showed no statistically significant differences (t =0.946,P =0.541).The amplitude of pseudophakic accommodation was negative correlated with pupil diameter (r =-0.651,P =0.024).Conclusions The movement of aspheric IOLs in eyes after cataract surgery is an important mechanism for attaining good near vision without additional correction,2-point haptic design IOLs provide better pseudophakic accommodation than that of Four-point haptic designs.
What problem does this paper attempt to address?