Experimental study of the wavefront aberrations of intraocular lenses in vitro
WANG Lu,TANG Xin,ZHAO Xing,王璐,汤欣,赵星,方志良
DOI: https://doi.org/10.3760/CMA.J.ISSN.1008-1801.2008.06.007
2008-01-01
Abstract:Objective To study the spherical aberrations(SA)of spherical and aspherical intraocular lenses(IOLs)in vitro.Methods A special optical metric system was set up to measure the wave-front aberrations of IOLs with a Hartmann-Shack aberrometer.The lenses were divided into 3 groups:CeeOn911A and Tecnis Z9001 lenses(group 1),AR40e and Tecnis ZA9003 lenses(group 2),and AcrySof Natural and AcrySof IQ lenses(group 3).IOL aberrations were measured continuously under 3 mm,4 mm,5 mm,6 mm and 7 mm apertures.The data was then statistically analyzed.Re-sults Spherical IOLs had a positive SA,aspherical IOLs had a negative SA,and significant intergroup differences were found:In the 3 mm aperture(t1=14.19,t2=12.56,t3=8.08,P<0.01),in the 4 mm aperture(t1=31.19,t2=11.79,t3=18.38,P<0.01),in the 5mm aperture(t1=23.02,t2=29.56,t3=41.39,P<0.01),in the 6 mm aperture(t1=60.77,t2=132.7,t3=57.42,P<0.01),and in the 7 mm aperture(t1=62.17,t2=52.43,t3=57.26,P<0.01).The total aberrations and SA for all of the IOLs increased with an increase in aperture size.There were positive correlations between the size of the aperture and total aberrations and spherical aberration(r = 0.9257,0.8274,0.7992,P <0.01).Conclusion Aspherical IOLs produce a negative SA.The lenses are able to compensate for the positive corneal SA and reduce ocular wavefront aberrations so vi-sual quality is improved.