A Clinical Study of Visual Function after Acri Smart 36A-5 Intraocular Lens Implantation

ZHAO Ke-ke,ZHAO Yun-e,WANG Zhao-yang,WANG Qin-mei
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2009.10.014
2009-01-01
Abstract:Objective To investigate the visual function after implantation of Acri Smart 36A-5 aspheric intraocular lens(IOL)in patients with age-related cataract.Methods We continuous selected 112 eyes of 90 age-related cataract patients requiring cataract surgery from Jun.2005 to Aug.2006 in our hospital.The patients were randomized to receive one of these three IOL typesd: Acri Smart 36A-5IOL(as group A,37 eyes),Teenis Z9001 IOL(as group B,39 eyes)or AR40e IOL(as group C,36 eyes).After 3 months,measure of refraction and best spectacle corrected visual acuity(BCVA),and contrast sensitivity with or without glare(CSV-1000)were examined.Results(1)There were not significantly different among 3 groups in BCVA in 3-month postoperatively(P=0.473).(2)Without glare(0.61ux),in the 6c/d,12c/d and 18c/d frequencies,there were significantly higher spatial contrast sensitivity in group A and group B than group C.With glare(151ux),in 6c/d,18c/d,there were statistically significant differences between group A or group B and group C.With glare(851ux),in 18c/d,there were statistically significant differences between group A or group B and group C.There were not significantly difference between group A and group B in any glare condition.Conclusion(1)Aspheric IOLs Acri Smart 36A-5 can increase the contrast sensitivity of some frequencies with or without glare after cataract surgery.(2)There is no difference between these two Aspheric IOLs in visual function.
What problem does this paper attempt to address?