2450 MHz and 915 MHz Microwave Ablation in Ex Vivo Bovine Livers:the Difference in Ablation Zone, Energy and Energy Conversion

Jiang Xiongying,Chen Dong,Ni Jiayan,Wang Weidong,Luo Jianghong,Chen Yaoting,Sun Hongliang,Xu Linfeng
DOI: https://doi.org/10.3877/cma.j.issn.2095-5782.2016.01.011
2016-01-01
Abstract:Objective To compare ablation zone, total deposited energy and energy conversion efifciency between 915 MHz non-cooled shaft and 2450 MHz cooled shaft microwave ablation in ex vivo bovine livers. Methods Experiment was carried out in ten fresh excised bovine livers weight 5.2-6.5 kg. The ten livers were divided into two groups, with Group A be 2450 MHz cooled shaft microwave ablation, and Group B be 915 MHz non-cooled shaft microwave ablation. In each group, different ablation durations (10, 20, 30, and 40 minutes) were evaluated. We compared coagulation diameters and volume, sphericity ratio, deposited energy, energy conversion efficiency in each ablation time.Results Compare to Group B, the mean coagulation volume and transverse diameter were signiifcantly greater with Group A and longitudinal diameter was lower (P<0.001). The sphericity ratio was closer to 1 with Group A. Total deposited energy with Group A was signiifcantly larger than in Group B. The value of total deposited energy with Group A was about two times with Group B in the same ablation duration. The energy conversion efficiency between the two groups was not statistically significantly differ for 10 minutes ablation but increased signiifcantly with Group B than with Group A from 20 minutes to 40 minutes (P<0.05 in all). Conclusions Due to more energy deposition, 2450 MHz cooled shaft microwave ablation can create substantially larger and more spherical ablation lesionsthan 915 MHz non-cooled shaft microwave ablation. But the energy conversion efifciency with 915 MHz non-cooled shaft microwave ablation is higher.
What problem does this paper attempt to address?