Comparison of Temperature Curve and Ablation Zone Between 915- and 2450-Mhz Cooled-Shaft Microwave Antenna: Results in Ex Vivo Porcine Livers.

Yuanyuan Sun,Zhigang Cheng,Lei Dong,Guoming Zhang,Yang Wang,Ping Liang
DOI: https://doi.org/10.1016/j.ejrad.2011.02.013
2011-01-01
Abstract:Objective: To compare temperature curve and ablation zone between 915- and 2450-MHz cooled-shaft microwave antenna in ex vivo porcine livers.Materials and methods: The 915- and 2450-MHz microwave ablation and thermal monitor system were used in this study. A total of 56 ablation zones and 280 temperature data were obtained in ex vivo porcine livers. The output powers were 50, 60, 70, and 80 W and the setting time was 600 s. The temperature curve of every temperature spot, the short- and long-axis diameters of the coagulation zones were recorded and measured.Results: At all four power output settings, the peak temperatures of every temperature spot had a tendency to increase accordingly as the MW output power was increased, and except for 5 mm away from the antenna, the peak temperatures for the 915 MHz cooled-shaft antenna were significantly higher than those for the 2450 MHz cooled-shaft antenna (p < 0.05). Meanwhile, the short- and long-axis diameters for the 915 MHz cooled-shaft antenna were significantly larger than those for the 2450 MHz cooled-shaft antenna (p < 0.05).Conclusion: The 915 MHz cooled-shaft antenna can yield a significantly larger ablation zone and achieve higher temperature in ablation zone than a 2450 MHz cooled-shaft antenna in ex vivo porcine livers. (C) 2011 Elsevier Ireland Ltd. All rights reserved.
What problem does this paper attempt to address?