Effect Analysis of Multi-Electrode Synchronous Radiofrequency Ablation and Conventional Single Electrode Overlapping Ablation

Yanyang Zhang,Yangkui Gu,Jinhua Huang,Fei Gao,Xiongying Jiang,Tianqi Zhang,Senmiao Huang
DOI: https://doi.org/10.3760/cma.j.issn.0376-2491.2015.43.013
2015-01-01
Abstract:OBJECTIVE:To explore the different effects in ex vivo bovine liver between multiple cool-tip internally cooled electrodes synchronous radiofrequency ablation via switching controller and conventional single electrode overlapping radiofrequency ablation. METHOD:The above two methods were used to ablate in ex vivo bovine liver respectively. Electrode exposed end was 3 cm or 4 cm in length. The morphological characteristics of lesions created by two methods were observed, the sizes of the ablated lesions were measured and compared. RESULT:The formation of lesions ablated by both two methods was ellipsoidal. The longitudinal diameter, transverse diameter, front-end diameter and volume of the lesions ablated 16 mins by multi-electrode synchronous radiofrequency ablation via switching controller were (7.12 ± 0.52) cm, (5.71 ± 0.47) cm, (1.77 ± 0.29) cm, (102 ± 20) cm³, respectively. All of them are larger than the lesions ablated 36 mins by conventional single electrode overlapping radiofrequency ablation (P<0.01). At the same 16 mins, multi-electrode synchronous radiofrequency ablation via switching controller created longer longitudinal diameter and large volume (P<0.01) with longer exposed end electrodes (4 cm vs 3 cm). But the transverse diameter, the front-end diameter had no statistical difference (P>0.05). CONCLUSION:Compared with the conventional single electrode overlapping radiofrequency ablation, multi-electrode synchronous radiofrequency ablation via switching controller create larger lesion in the shorter time. It may provide a new attempt for the clinical treatment of large hepatocellular carcinoma.
What problem does this paper attempt to address?