Radiofrequency Ablation in Normal Swine Liver: in Vivo Experimental Study

FENG Yun,TONG Xiao-qiang,YANG Min,ZOU Qiang,LV Yong-xing,ZOU Ying-hua
DOI: https://doi.org/10.3969/j.issn.1672-8475.2005.03.016
2005-01-01
Abstract:Objective To investigate the radiological and pathological features of ablated lesions obtained by the expandable RF system, and the influence of saline injection during RF ablation on lesion size. Methods In vivo radiofrequency ablation (RFA) was processed on 5 normal swine with the 3rd gear of electrode size. The ablation temperature was 100℃ while the treatment time was 20 minutes. RFA with saline injection was performed in 6 lesions and without in 5, resulting in 11 lesions as a whole. RFA lesions were examined by triple-phase contrast-enhanced CT immediately after the procedure and several weeks later on, respectively, and all swine were sacrificed finally for pathological evaluation. Results The mean diameter of RFA lesions obtained with temperature of 100℃ and 3rd gear of electrode's size was 3.16 cm, while that with saline injection was 4.08 cm on post-procedural triple-phase CT scan. Preoperative injection of saline could significantly increase RFA lesion size compared with that of non-injection group (P0.05), whether on post-procedural CT scan or follow-up. On CT scans during follow-up, the RFA lesions shrunk in various extent correlating with time and fibrous repair. Coagulation necrosis was observed in HE stained slices of RFA lesions which manifested as non-enhanced areas on enhanced CT. Conclusion The used radiofrequency system can result in coagulative necrosis reliably and definitely, and injection of saline may increase the size of coagulated area.
What problem does this paper attempt to address?