Determination of Two Anomers of Tribenoside by HPLC

SONG Li,Zhang Dan
DOI: https://doi.org/10.3321/j.issn:1001-2494.2005.14.020
2005-01-01
Abstract:OBJECTIVE To establish a HPLC for the determination of two anomers of tribenoside. METHODS The determination was performed on a Hypersil C18 column (4.6 mm×250 mm, 10 μm) with UV detection at 258 nm, and the mobile phase consisted of methyl alcohol and water(82:18). The flow rale was 1.0 mL·min-1.RESULTS The calibration curves were linear ( r=0.999 9) within the range of 25.6-512.0 mg·L-1 for α-anomer.The precision of injection for the high,middle and low concentration was blow 0.79% ( n=5),the detection limit was 0.15 μg(S/N = 3) .The precision of repeatability RSD was 0.56% (n=6), the intermedia precision RSD was 0.65% (n=5). The calibration curves were linear (r =0.999 8) within the range of 75.2-1128.0 mg·L-1 for β-anomer.The precision RSD of injection for the high, middle and low concentration was below 0.64% (n=5), the detection limit was 0.05 μg(S/N = 3), the precision of repeatabil-ity RSD was 0.17% (n=6), the intermedia precision RSD was 0.36% ( n=5). CONCLUSION This method is simple, accurate and the result of determination is good. This method is suitable lor the quality control of tribenoside and its preparation.
What problem does this paper attempt to address?